Screening drugs of abuse by LC/MS

Keysight Technologies Australia Pty Ltd
Sunday, 08 May, 2005


High throughput screening of drugs of abuse is performed at St Olav Hospital by LC/MS. Over a million analyses per year are now made. Typically done by immunoassay, this overview describes the procedures for using this highly selective and quantitative LC/MS methodology. In addition, the advantages of using LC/MS (lower cut-offs, no false positives, etc) are discussed.

Substance abuse is increasing in almost every country, and is associated with criminality, health hazards, accidents and fatalities. Communities, therefore, face a significant resource investment in their efforts to reduce or stop this abuse. Most substance abuse is illegal. Symptoms and signs are unspecific, and substance abuse is difficult to identify outside the boundary of clinical criteria. This, combined with the fact that substance abuse under many circumstances leads to legal actions, makes it mandatory to be able to identify and quantify substances of abuse in biological material. Such methods were developed and are applicable for almost any possible biological matrix.

Traditionally, screening is done by immunology, which is fast and simple, but can be expensive (reagent costs), and normally determines groups of compounds, not specific analytes. Due to its lack of specificity, very often positives must be confirmed, normally by gas chromatography/mass spectrometry (GC/MS). In drug screening, immunology gives a result as 'positive' or 'negative', with reference to a certain pre-determined cut-off level. Cut-off values for immunoassays are fixed due to optimisation of quantity, and tend to be relatively high to avoid bias from interferences. As a result, this gives a high number of false negatives that may have clinical and societal consequences.

A drug screen by liquid chromatography/mass spectrometry (LC/MS) gives a quantitative determination of specific analytes, with known accuracy and precision, within a range of concentrations from 50-100,000 ng/mL. This allows variable cut-off levels for different purposes within the calibrated range. An argument can be made that if a compound is not included in the LC/MS screen it will be missed, and that immunoassay will give a positive in that case because it is a general screen. However, confirmation will show it to be a false positive because the GC/MS confirmation is also a targeted list. Screening by LC/MS is a new approach compared to immunoassay. LC/MS is also fast, but provides results for specific compounds, not groups. This is important, for example, where the patient is legally prescribed one benzodiazepine but may abuse a second benzodiazepine. There is no way to account for this with immunoassay; however, intake of other 'non-prescribed' benzodiazepines is easily detected by LC/MS screening. A similar argument can be applied to amphetamines and other groups of drugs. As an example, LC/MS screening of amphetamines can differentiate between the following analytes: amphetamine, methamphetamine, methylenedioxymethamphetamine (MDMA or Ecstasy), methylenediox-amphetamine (MDA), and ephedrine.

LC/MS methods for 'new' drugs on the street can be quickly developed, validated and implemented into the assay within a few days. In the example of amphetamines, other related drugs such as cathinone can be easily and quickly added to the screen.

This is not the case for immunoassay, where development of kits for new analytes is a challenging and time-consuming procedure. LC/MS is flexible, reliable and highly sensitive (low nanogram range). As part of its flexibility, note that systems used for other purposes, such as therapeutic drug monitoring (TDM), can also be used for drugs of abuse screening and vice versa. This system flexibility and versatility is an im-portant feature of the platform and is important both for logistics and maintenance.

LC/MS at St Olav Hospital

This overview describes the successful use of LC/MS systems at St Olav Hospital in routine service doing high-volume drug screens from 1998 to the present. Figure 1 shows the increase in the number of analyses performed each year during the period of 1996 to 2003. The first LC/MS was put into service in 1998 and the methodology was fully employed by 1999. The number of analyses for 2004 approached 1,000,000. Note that because of the graph's scale, the increase in serum analyses cannot be read, but the number of serum determinations is increasing. Serum analyses are approaching 60,000 for this year. Each DOA analysis represents a determination equivalent to an immunoassay for a group of drugs (benzo-diapenes, amphetamines, etc). The actual LC/MS analysis determines specific compounds, and if charted by the compounds analysed, the total number of analyses would be much higher. LC/MS screening is now performed as a routine service in a restrict-ed area in compliance with national and in-ternational guidelines using quality control systems securing all aspects of sample handling, preparation, analysis and reporting.

Methodology

The LC/MS platform is used for a wide variety of samples ranging from medical treatment of abuse, legal actions/negative sanctions, forensics, and clinical toxicology. Several types of these samples must be confirmed. Because GC/MS is still the accepted 'gold standard' confirmation technique for legal action, this is the methodology used here.

However, the use of LC/MS screening strongly reduces the number of GC/MS confirmations, a fact that saves both time and money. Comparison of GC/MS and LC/MS results show close to 100% accordance, which means no false positives. In the future, a high-throughput technology such as liquid chromatography/tandem mass spectrometry (LC/MS/MS) in full scan mode, as obtained by ion trap technology, may demonstrate the potential for performing fast confirmation. In combination with LC/MS screening, such a technique would make possible screening and confirmation in less than an hour for single samples, and within a few hours for a larger series of samples.

The systems for DOA screening and TDM use the Agilent 1100 LC/MSD quadruples. Presently, 24 instruments are used for these activities. All instruments (both for DOA and TDM) are equipped identically with four mobile phase constituents, using a quaternary system with methanol, acetonitrile, ammonium acetate, and formic acid. For DOA, only two columns are needed, a short C18 and a short CN. This simple strategy gives unique flexibility between instruments and very efficient backup capacity. Finally, the simplified inventory of mobile phases and columns makes fast method development easier.

Amphetamines as an example

As an example, amphetamines are determined with a short CN column with an isocratic mobile phase (ammonium acetate and acetonitrile).

Figure 2 shows amphetamine, methamphetamine, MDA, MDMA and ephedrine with d3-amphetamine as the internal standard (ISTD). Target ions and qualifiers are used, and the mass spectrometer is operated using electrospray ionisation (ESI). The qualifier ions are obtained by collision-induced dissociation (CID) in the ion transport region of the atmospheric pressure ionisation (API) interface, commonly known as 'up-front or in-source CID'. Note that little chromatographic separation is achieved with the fast run time. However, the single quadrupole mass spectrometer provides sufficient selectivity to separate each compound with quantitative accuracy.

The qualifier ions provide additional selectivity to assure confidence in the determination of each compound. With liquid-liquid extraction of the urine samples, sufficient clean-up is achieved for the analysis.

Even though fast chromatography is used, there is sufficient retention for each of the analytes to be moved from the void of the column.

For complete screening of all categories of drugs of abuse, both ESI and APCI (atmospheric pressure chemical ionisation) must be employed. An example of a complete group of DOA compounds best analysed by APCI is the benzodiazepines. Some of the compounds in this category do respond well to ESI, but others do not. All do respond well to APCI. In this method, a short C18 column with gradient conditions using a mix of methanol, formic acid, and ammonium acetate, provides the best results in a relatively short time.

Quality assurance

To obtain the highest quality results, processes must be in place to assure that the instruments are running properly and that all extractions and analyses are done correctly. This assurance is provided by both internal (prepared in the laboratory) and external (obtained by sources outside the laboratory) quality control samples. Every batch of samples analysed contains the internal quality control samples at concentrations covering the range of concern for the analytes.

Table 1 shows typical results obtained for these QC samples. These QC results indicate not only the quality of the determination of each specific target compound, but their concentration as well.

  QC50 QC100 QC500 QC2000
Amphetamine 48 97 517 2039
Methamphetamine 58 109 533 2049
MDMA 59 112 537 2029
MDA 50 98 517 2140
Ephedrine 56 107 512 2067
Morphine 53 105 526 1993
Codeine 52 108 511 2102
Methadone 53 104 507 2018
Benzoylecognine 59 112 503 2119
Phencyclidine (1/10) 5 10 50 224
Table 1: Internal QC results for some DOA.

Conclusions

The laboratory at St Olav Hospital routinely analysed 800,000 DOA urine samples and 30,000 TDM serum samples in 2003, using 24 LC/MS systems. This year the number is approaching 1 million analyses, taking into consideration that, for example, the amphetamine group (with five analytes) is only counted as a single analysis. This is also the case for the benzodiazepines (six analytes) as well as the opiates (four analyts). The accounting scheme is mainly for administrative reasons and for easier comparison with immunology-based laboratories.

Twelve systems are set up using ESI and 12 systems using APCI and the instrument configurations are flexible enough to perform both DOA analysis and TDM analyses. The DOA screens include amphetamines, benzodiazepines, opiates, metha- done, buprenorphine, PCP, cocaine and its metabolite, barbiturates, and others.

The procedures and the instrumentation briefly described here allow this laboratory to perform these analyses both in a cost-effective way and with the highest quality results possible.

In addition, the laboratory uses 10 GC/MS instruments, both for confirmation and unknown compound identification. It should be emphasised that this combination of LC/MS and GC/MS instruments comprise a very strong analytical platform, especially for forensic and clinical toxicology. The laboratory performs several thousand analyses per year for these categories. Biological concentrations of specific drugs with secure identification and fast results is of great importance for the clinicians to make assessments toward the dose and clinical state of a patient. This analytical platform for these determinations requires a significant initial capital investment, but the return in both efficiency and medical quality of the results provide justifiable benefits.

Kolbjorn Zahlsen and Trond Aamo, Department of Clinical Pharmacology, St Olav Hospital, University Hospital of Trondheim, Trondheim, Norway - Jerry Zweigenbaum, Agilent Technologies, Inc

Related Articles

Mast cell test simplifies the diagnosis of food allergies

In the Hoxb8 mast cell activation test (Hoxb8 MAT), mast cells grown in the laboratory are...

A science-based solution for tackling lake health

Experts urge moving away from short-term 'fixes' that only worsen a lake's condition...

Climate report warns of perilous times ahead

Scientists have warned that the Earth is stepping into a critical and unpredictable new phase of...


  • All content Copyright © 2024 Westwick-Farrow Pty Ltd